Yokogawa
YTA610 Temperature Transmitter
Dual-input RTD/thermocouple transmitter with sensor backup and drift detection.
- Inputs
- Dual RTD or thermocouple
- Accuracy
- +/-0.1 C (Pt100)
- Diagnostics
- Sensor backup and drift detection
End-to-end OT partner: supply, integration, security and lifecycle support.
Yokogawa · Pressure Transmitters
High-stability DP transmitter for flow, level and pressure duty in process plants.
The reference DP transmitter for orifice flow and hydrostatic level. Its long-term stability specification is the reason it appears in so many process plants: fewer recalibration visits over the asset life, which usually outweighs the purchase premium on a ten-year view.
| Measurement | Differential pressure |
|---|---|
| Span limits | 0.1 kPa to 14 MPa across capsule codes |
| Reference accuracy | +/-0.055 % of span |
| Long-term stability | +/-0.1 % of URL over 10 years |
| Output signal | 4-20 mA DC with HART 7 |
| Wetted materials | 316L SST, Hastelloy options |
| Process connection | 1/2 NPT, flanged options |
| Ambient temperature | -40 to +85 C |
| Hazardous area | ATEX, IECEx intrinsically safe |
| Enclosure | IP67 |
Specifications are indicative and subject to variant. The manufacturer datasheet governs in all cases — ask us for the current revision against your part number.
Datasheets, dimensional drawings, wiring schematics and declaration certificates are issued on request against your specific part number and variant. Request documentation.
Accessories and companion equipment that usually appear on the same bill of materials.
Yokogawa
Dual-input RTD/thermocouple transmitter with sensor backup and drift detection.
Yokogawa
Electromagnetic flow meter for conductive liquids, with dual-frequency excitation.